Automatic external defibrillator: Welch Allyn AED 10
Device recall due to quality and performance problems. Devices can fail suddenly and prevent defibrillation of a patient
Welch Allen have instigated a field recall for some of their Defibrillators which are branded as
MRL JumpStart
Automatic external defibrillatorWelch Allyn AED 10 was previously branded as MRL LifeQuest JumpStart in the UK.
The AED 10 is a portable, semi-automatic external defibrillator used in hospitals, ambulances and public places.
Depending upon the date of manufacture, units may have one or more of the following problems as described briefly below and in detail in the manufacturer's Field Safety Notice.- Low energy shock
- Electromagnetic noise interference
- Unexpected shutdown during use
- Blown fuse resulting in device failure
- Loss of voice prompts
- Shut down in cold temperatures
The serial number is located on the top left of the base of the unit.
Further information and reporting procedures can bee seen using the link below.Welch Allen defibrillator recall